SMOKE TEST

Information

For the control of the formation of vortices that could compromise the "containment" and disturbances caused by air currents in the Laboratory.

CONTROL OPENING FORCE

CONTROLLO FORZA DI APERTURA

According to the EN 14175 standard, the force to be used to move the sliding rail must not exceed 30 Nw if manually lifted and the movement speed for motorized up and down must not exceed 0.5 m / s.

CONTAINMENT TEST OF FUME CUPBOARD

CONTAINMENT TEST OF FUME CUPBOARD

For the control of the formation of vortices that could compromise the "containment" and disturbances caused by air currents in the Laboratory.
• SF6 tracer gas metering system
• SF6 tracer gas flow control system
• Mixture of SF6 gases with a known concentration
• Point-on-point system for the SF6 gas
• Gas concentrator coming from the sensors
• SF6 trace analyzer
• Translator panel tests robustness.